Cross-Sectional Computed Tomographic Assessment Improves Accuracy of Aortic Annular Sizing for Transcatheter Aortic Valve Replacement and Reduces the Incidence of Paravalvular Aortic Regurgitation

OBJECTIVES: In an effort to define the gold standard for annular sizing for transcatheter aortic valve replacement (TAVR), we sought to critically analyze and compare the predictive value of multiple measures of the aortic annulus for post-TAVR paravalvular (PV) regurgitation and then assess the impact of a novel cross-sectional computed tomographic (CT) approach toannular sizing. Recent studies have shown clear discrepancies between conventional 2-dimensional (2D) echocardiographic and CT measurements. In terms of aortic annular measurement for TAVR, such findings have lacked the outcome analysis required to inform clinical practice.

METHODS: The discriminatory value of multiple CT annular measures for post-TAVR PV aortic regurgitation was compared with 2D echocardiographic measures. TAVR outcomes with device selection according to aortic annular sizing using a traditional 2D transesophageal echocardiography-guided or a novel CT-guided approach were also studied.

RESULTS: In receiver-operating characteristic models, cross-sectional CT parameters had the highest discriminatory value for post-TAVR PV regurgitation: This was with the area under the curve for [maximal cross-sectional diameter minus prosthesis size] of 0.82 (95% confidence interval: 0.69 to 0.94; p < 0.001) and that for [circumference-derived cross-sectional diameter minus prosthesis size] of 0.81 (95% confidence interval: 0.7 to 0.94; p < 0.001). In contrast, traditional echocardiographic measures were nondiscriminatory in relation to post-TAVR PV aortic regurgitation. The prospective application of a CT-guided annular sizing approach resulted in less PV aortic regurgitation of grade worse than mild after TAVR (7.5% vs. 21.9%; p = 0.045).

CONCLUSIONS: Our data lend strong support to 3-dimensional cross-sectional measures, using CT as the new gold standard foraortic annular evaluation for TAVR with the Edwards SAPIEN device.

PMID: 22365424

Posted in Magnetic Resonance Imaging and tagged , , , , .

2 Comments

  1. See also:

    3-Dimensional Aortic Annular Assessment by Multidetector Computed Tomography Predicts Moderate or Severe Paravalvular Regurgitation After Transcatheter Aortic Valve Replacement A Multicenter Retrospective Analysis.
    Willson AB, Webb JG, Labounty TM, Achenbach S, Moss R, Wheeler M, Thompson C, Min JK, Gurvitch R, Norgaard BL, Hague CJ, Toggweiler S, Binder R, Freeman M, Poulter R, Poulsen S, Wood DA, Leipsic J.
    J Am Coll Cardiol. 2012 Feb 7.
    PMID: 22365423 http://www.ncbi.nlm.nih.gov/pubmed/22365423

  2. See recent MRI reference:

    Comparison of magnetic resonance imaging of aortic valve stenosis and aortic root to multimodality imaging for selection of transcatheter aortic valve implantation candidates.
    Paelinck BP, Van Herck PL, Rodrigus I, Claeys MJ, Laborde JC, Parizel PM, Vrints CJ, Bosmans JM.
    Am J Cardiol. 2011 Jul 1;108(1):92-8.
    PMID: 21529729 http://www.ncbi.nlm.nih.gov/pubmed/21529729