Effect of Reduced Radiation Exposure and Iterative Reconstruction on Detection of Low-Contrast Low-Attenuation Lesions in an Anthropomorphic Liver Phantom: An 18-Reader Study

OBJECTIVES: To measure the effect of reduced radiation exposure on low-contrast low-attenuation liver lesion detection in an anthropomorphic abdominal phantom by using filtered back projection (FBP) and sinogram-affirmed iterative reconstruction.

METHODS: Eighteen radiologists blinded to phantom and study design interpreted randomized image data sets that contained 36 spherical simulated liver lesions of three sizes and three attenuation differences (5-mm diameter: 12, 18, and 24 HU less than the 90-HU background attenuation of the simulated liver insert; 10- and 15-mm diameter: 6, 12, and 18 HU less than the 90-HU background attenuation) scanned with four discrete exposure settings and reconstructed by using FBP and sinogram-affirmed iterative reconstruction. Response assessment included region-level lesion presence or absence on a five-point diagnostic confidence scale. Statistical evaluation included multireader multicase receiver operating characteristic curve analysis, with nonparametric methods and noninferiority analysis at a margin of -0.10.

RESULTS: Pooled accuracy at 75% exposure for both FBP and sinogram-affirmed iterative reconstruction was noninferior to 100% exposure (P = .002 and P < .001, respectively). Subsequent exposure reductions resulted in a significant decrease in accuracy. When the smallest (5-mm-diameter) lesions were excluded from analysis, sinogram-affirmed iterative reconstruction was superior to FBP at 100% exposure (P = .011), and sinogram-affirmed iterative reconstruction at 25% and 50% exposure reduction was noninferior to FBP at 100% exposure (P ≤ .013). Reader confidence was greater with sinogram-affirmed iterative reconstruction than with FBP for 10- and 15-mm lesions (2.94 vs 2.76 and 3.62 vs 3.52, respectively).

CONCLUSIONS: In this low-contrast low-attenuation liver lesion model, a 25% exposure reduction maintained noninferior diagnostic accuracy. However, detection was inferior with each subsequent exposure reduction, regardless of reconstruction method. Sinogram-affirmed iterative reconstruction and FBP performed equally well at modest exposure reduction (25%-50%). Readers had higher confidence levels with sinogram-affirmed iterative reconstruction for the 10- and 15-mm lesions. John Lynch Authentic Jersey

PMID: 24620913

Posted in Computed Tomography and tagged , , , , .

3 Comments

  1. See also:

    Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver.
    Baker ME, Dong F, Primak A, Obuchowski NA, Einstein D, Gandhi N, Herts BR, Purysko A, Remer E, Vachhani N.
    AJR Am J Roentgenol. 2012 Jul;199(1):8-18.
    PMID: 22733888. http://www.ncbi.nlm.nih.gov/pubmed/22733888

  2. It is a privilege to have our article displayed on thepreparedminds.com. Thank you Dr. Schoenhagen and team.

    In this article, we have highlighted the need for radiation dose reduction to be optimized to the diagnostic task under consideration. While subjective perception of image quality and objective measures are important, the ultimate test is diagnostic performance of readers at reduced radiation exposures when compared with conventional technology at standard radiation exposures.